R&S FSP
Remote Control – Description of Commands
DISPlay Subsystem
6.108
Operating Manual 1164.4556.12 - 06
DISPlay:FORMat
SINGle | SPLit
This command switches the measurement result display between FULL SCREEN and SPLIT
SCREEN. The coupling of settings between screen A and screen B can be selected with the command
INSTrument:COUPle
.
In full-screen display the active measurement window can be selected with
DISPlay:WINDow<1|2>:
SELect
.
Example:
"DISP:FORM SPL"
'Switches the display to 2 measurement windows.
Characteristics:
*RST value: SINGle
SCPI: device-specific
Mode:
A
DISPlay:LOGO
ON | OFF
This command switches the company logo on the screen on or off.
Example:
"DISP:LOGO OFF"
Characteristics:
*RST value: ON
SCPI: device-specific
Mode:
A
DISPlay:PSAVe:HOLDoff
1 to 60
This command sets the hold off time for the power-save mode of the display. The available value range
is 1 to 60 minutes, the resolution 1 minute. The entry is dimensionless.
Example:
"DISP:PSAV:HOLD 30"
Characteristics:
*RST value: 15
SCPI: device-specific
Mode:
A
DISPlay:PSAVe[:STATe]
ON | OFF
This command switches on or off the power-save mode of the display. With the power-save mode
activated the display including backlight is completely switched off after the elapse of the response
time (see command
DISPlay:PSAVe:HOLDoff
).
Example:
"DISP:PSAVe ON"
'Switches on the power-save mode.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
This mode is recommended for preserving the display especially if the instrument
is exclusively operated via remote control.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...