Operating Manual 1164.4556.12 - 06
6.227
R&S FSP
Remote Control – Description of Commands
SYSTem Subsystem
SYSTem:DISPlay:UPDate
ON | OFF
This command switches on or off the update of all display elements during remote control.
Example:
"SYST:DISP:UPD ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
SYSTem:ERRor?
This command queries the earliest entry in the error queue, and deletes it after the readout.
Positive error numbers indicate device-specific errors, negative error numbers are error messages
defined by SCPI (cf. chapter
). If the error queue is empty, the error number 0, "no
error", is returned. This command is identical with the command
STATus:QUEue:NEXT?
.
This command is a query and therefore has no *RST value.
Example:
"SYST:ERR?"
Characteristics:
*RST value: –
SCPI: conform
Mode:
A
SYSTem:ERRor:CLEar:ALL
This command deletes all entries in the table SYSTEM MESSAGES.
This command is an event and therefore has no query and no *RST value.
Example:
"SYST:ERR:CLE:ALL?"
Characteristics:
*RST value: –
SCPI: device-specific
Mode:
A
SYSTem:ERRor:LIST?
This command reads all system messages and returns a list of comma separated strings. Each string
corresponds to an entry in the table SYSTEM MESSAGES.
If the error list is empty, an empty string "" will be returned.
This command is a query and therefore has no *RST value.
Example:
"SYST:ERR:LIST?"
Characteristics:
*RST value: –
SCPI: device-specific
Mode:
A
The best performance is obtained when the display output is switched off during
remote control.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...