R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.162
Operating Manual 1164.4556.12 - 06
[SENSe<1|2>:]ESPectrum:RANGe<1...20>[:FREQuency]:STOP
<numeric_value>
This command sets the relative stop frequency of a range in the Spectrum Emission Mask
measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"ESP:RANG2:STOP 2.52MHZ"
'stop frequency of range 2 to +2.52MHz
Characteristics:
*RST value:-
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:INPut:ATTenuation
<numeric_value>
This command defines the RF attenuation of a range in the Spectrum Emission Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"
ESP:RANG2:INP:ATT 30db"
'RF attenuation of range 2 to 30 dB
Characteristics:
*RST value:-
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:INPut:ATTenuation:AUTO
ON | OFF
This command switches the attenuation mode of a range in the Spectrum Emission Mask
measurement to AUTO (ON) or MANUAL (OFF).
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"ESP:RANG2:INP:ATT:AUTO ON"
'attenuation mode is AUTO for range 2
Characteristics:
*RST value: ON
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:INPut:GAIN:STATe
ON | OFF
This command switches the preamplifier of a range in the Spectrum Emission Mask measurement on
or off.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"ESP:RANG2:INP:GAIN:STAT ON"
' activates the preamplifier for range 2
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...