Press the
VIDEO BW MANUAL
softkey and enter
10 kHz
.
The RF signal can be more clearly distinguished from noise.
Fig. 2-3
RF sinewave signal with low S/N ratio if a smaller video bandwidth is
selected.
7. By reducing the resolution bandwidth by a factor of 10, the noise is reduced
by 10 dB.
➢
Press the
RES BW MANUAL
softkey and enter
300 kHz
.
The displayed noise is reduced by approx. 10 dB. The signal, therefore,
emerges from noise by about 10 dB. Compared to the previous setting, the
video bandwidth has remained the same, i.e. it has increased relative to the
smaller resolution bandwidth. The averaging effect is, therefore, reduced by
the video bandwidth. The trace will be noisier.
Fig. 2-4
Reference signal at a smaller resolution bandwidth
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...