Operating Manual 1164.4556.12 - 06
6.67
R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:LIST?
This command reads out the list of harmonics. The first value is the absolute power of the first
harmonic in the unit set via UNIT. The other values are relative to the carrier signal and are output
in dB.
The numeric values are output, separated by commas. The number corresponds to that of the
harmonics to be measured, which are defined with
CALC:MARK:FUNC:HARM:NHARM
.
A complete sweep with synchronization to sweep end should be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in the
single-sweep mode.
The function is independent of the marker selection, i.e. the suffix
<1|2>
or
<1...4>
of
CALCulate
or
MARKer
is irrelevant.
Example:
"INIT:CONT OFF"
’Switches to single-sweep mode.
"CALC:MARK:FUNC:HARM:NHARM 3"
'Sets the number of harmonics to 3.
"CALC:MARK:FUNC:HARM ON"
’Switches on the measurement of harmonics.
"INIT;*WAI"
’Starts a sweep and waits for the end.
"CALC:MARK:FUNC:HARM:LIST?"
’Outputs the three measured harmonics, separated by commas.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A-F, A-T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:NHARmonics
1 to 26
This command defines the number of harmonics of a carrier signal to be measured. The function is
independent of the marker selection, i.e. the suffix
<1|2>
or
<1...4>
of
CALCulate
or
MARKer
is
irrelevant.
Example:
"CALC:MARK:FUNC:HARM:NHARM 3"
'Sets the number of harmonics to be measured to 3.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A-F
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...