Operating Manual 1164.4556.12 - 06
6.193
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]POWer:ACHannel:FILTer:STATe:CHANnel<1...12>
ON | OFF
This command switches the weigthing filter state for the TX channel on or off. It is only available if Fast
ACP is not active and no predefine standard is selected.
This command changes the setting for the TX channel. The numeric suffix at CHANnel is ignored.
Example:
"POW:ACH:FILT:STAT:CHAN ON"
'activates the weighting filter
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]POWer:ACHannel:MODE
ABSolute | RELative
This command toggles between absolute and relative adjacent channel measurement.
For the relative measurement the reference value is set to the currently measured channel power by
command
SENSe:POWer:ACHannel:REFerence:AUTO ONCE
.
The command is only available in the frequency domain (span > 0).
Example:
"POW:ACH:MODE REL"
Characteristics:
*RST value: ABSolute
SCPI: device-specific
Mode:
A-F
[SENSe<1|2>:]POWer:ACHannel:PRESet
ACPower | CPOWer | MCACpower | OBANdwidth |
OBWidth | CN | CN0
This command adjusts the frequency span, the measurement bandwidths and the detector as required
for the number of channels, the channel bandwidths and the channel spacings selected in the active
power measurement. If necessary, adjacent channel power measurement is switched on prior to the
adjustment.
To obtain valid results, a complete sweep with synchronization to the end of the sweep must be
performed after the adjustment. Synchronization is possible only in the single-sweep mode.
The result is queried with the command
CALCulate:MARKer:FUNCtion:POWer:RESult?
.
The command is available only for measurements in the frequency domain (span > 0).
Example:
"POW:ACH:PRES ACP"
'Sets the frequency span, the measurement bandwidths and the detector as
required for the ACP measurement in screen A.
"INIT:CONT OFF"
'Switches over to single-sweep mode.
"INIT;*WAI"
'Starts a sweep and waits for the end of the sweep.
"CALC:MARK:FUNC:POW:RES? ACP"
'Queries the result of the adjacent channel power measurement.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-F
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...