Operating Manual 1164.4556.12 - 06
6.127
R&S FSP
Remote Control – Description of Commands
INSTrument Subsystem
6.13 INSTrument Subsystem
The INSTrument subsystem selects the operating mode of the unit either via text parameters or fixed
numbers. Only operating modes available for the base unit and the models and the options described in
this manual (see
“Documentation Overview” on page 4.2
) are listed. For details on the other operating
modes refer to the corresponding separate manuals.
INSTrument:COUPle
NONE | RLEVel | CF_B | CF_A
In analyzer mode, this command selects the parameter coupling between the two measurement
windows screen A and B.
Parameter:
NONE: No coupling. The two measurement windows are operated like two
independent "virtual" devices.
RLEVel: The reference levels of the two measurement windows are coupled.
CF_B: The center frequency of screen B is coupled to the frequency of marker 1 in
screen A.
CF_A: The center frequency of screen A is coupled to the frequency of marker 1 in
screen B.
Example:
"INST:COUP NONE"
'Switches off the coupling of measurement windows. This leads to two independent
"virtual" devices.
Characteristics:
*RST value: NONE
SCPI: device-specific
Mode:
A
INSTrument:NSELect
<numeric value>
This command enables you to switch between the modes by using numbers.
Parameter:
1: Analyzer mode
3: FM demodulator mode
12: BLUETOOTH
Example:
"INST:NSEL 1"
'Switches to the analyzer mo
de
.
Characteristics:
*RST value: 1
SCPI: conform
Mode:
A
A changeover to a mode is only possible if the corresponding option is installed.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...