Operating Manual 1164.4556.12 - 06
4.275
R&S FSP
Instrument Functions
LO/IF ports for external mixers - Option R&S FSP-B21
Data and type designation of the mixer can be entered into the header of the
table.The columns contain the frequencies and the corresponding conversion
losses.
Name
- File name
The name of the table is identical with the name of the file (without extension) in
which the table is stored. The entry in this field is mandatory. The name may consist
of a maximum of 10 characters and ends with the harmonic used, e.g. Mixer_2. The
.ACL
extension will automatically be appended at storing.
Remote command:
CORR:CVL:SEL 'LOSS_T_4'
Mixer
- Type designation of mixer
The information in the
Mixer
field shows the user at a glance to which mixer a file
belongs. Entries in this field are optional.
Remote command:
CORR:CVL:MIX 'FS_Z60'
Band
- Designation of waveguide band
A waveguide band can be selected from a list. This field is mandatory.
Remote command:
CORR:CVL:BAND E
S/N
- Serial number
Serial number of the mixer. Entries in this field are optional.
Remote command:
CORR:CVL:SNUM '123.4567'
Harmonic#
- Order of harmonic
Order of harmonic used for conversion. For the conversion loss is valid for one har-
monic only, entries in this field are mandatory.
Remote command:
CORR:CVL:HARM 4
Ports
- Two- or three-port mixer
A list is displayed from which the number of ports (two or three) of the mixer used
can be selected. The entry in this field is mandatory.
Remote command:
CORR:CVL:PORT 3
Bias
- Bias setting
The bias current required for the mixer has to be entered in this field. The setting
range is between -10 mA and +10 mA.
The entry in this field is mandatory.
The current corresponds to the short-circuit current. The actual bias current is
lower because of the forward voltage of the mixer diode(s).
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...