R&S FSP
Remote Control – Description of Commands
TRIGger Subsystem
6.250
Operating Manual 1164.4556.12 - 06
Example:
"TRIG:SOUR EXT"
'Selects the external trigger input as source of the trigger signal
Characteristics:
*RST value: IMMediate
SCPI: conform
Mode:
A
This command is only available from firmware version 1.20 or higher.
TRIGger<1|2>[:SEQuence]:VIDeo:FIELd:SELect
ALL | ODD | EVEN
With active TV trigger this command activates triggering at the vertical sync signal.
The measurement is triggered on both fields with selection ALL, on odd fields with selection ODD and
on even fields with selection EVEN.
Example:
"TRIG:VID:FIEL:SEL ALL"
Characteristics:
*RST value:
SCPI: conform
Mode:
A-T
This command is only available with option R&S FSP-B6 (TV and RF Trigger).
TRIGger<1|2>[:SEQuence]:VIDeo:FORMat:LPFRame
525 | 625
This command defines the line system in use (525 or 625 lines) with active TV trigger.
Example:
"TRIG:VID:FORM:LPFR 525"
Characteristics:
*RST value: --
SCPI: conform
Mode:
A-T
This command is only available with option R&S FSP-B6 (TV and RF Trigger).
TRIGger<1|2>[:SEQuence]:VIDeo:LINE:NUMBer
<numeric_value>
With active TV trigger this command activates triggering at the horizontal sync signal of the indicated
line number.
Example:
"TRIG:VID:LINE:NUMB 17"
Characteristics:
*RST value:
SCPI: conform
Mode:
A-T
This command is only available with option R&S FSP-B6 (TV and RF Trigger).
IFPower
The next measurement is triggered by the detection of a signal at
the instrument IF (10 MHz bandwidth)
RFPower
The next measurement is triggered by the detection of a signal at
the instrument RF (80 MHz bandwidth)
TV
The next measurement is triggered by the detection of a TV signal
according to the settings of the
TRIGger:SEQuence:VIDeo
-
subsystem (not available for FM demodulator mode)
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...