[SENSe<1|2>:]AVERage[:STATe<1...3>]
ON | OFF
This command switches on or off the average calculation for the selected trace (STATe<1...3>) in the
selected measurement window.
Example:
"AVER OFF"
'Switches off the average calculation for trace 1 in screen A.
"SENS2:AVER:STAT3 ON"
'Switches on the average calculation for trace 3 in screen B.
Characteristics: *RST value:
OFF
SCPI: conform
Mode:
A
[SENSe<1|2>:]AVERage:TYPE
VIDeo | LINear
This command selects the type of average function. If VIDeo is selected, the logarithmic power is
averaged and, if LINear is selected, the power values are averaged before they are converted to
logarithmic values.
The type of average calculation is equally set for all traces in one measurement window.
Example:
"AVER:TYPE LIN"
'Switches screen A to linear average calculation.
Characteristics: *RST value:
VIDeo
SCPI: device-specific
Mode:
A
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...