Operating Manual 1164.4556.12 - 06
4.11
R&S FSP
Instrument Functions
Analyzer Mode
= MARKER
The
= MARKER
softkey sets the step size coupling to MANUAL and the step size to
the value of the marker. This function is especially useful during measurements of
the signal harmonic content at the marker position because by entering the center
frequency each stroke of the
STEP
key selects the center frequency of another har-
monic.
Remote command:
--
MANUAL
The
MANUAL
softkey activates the window for entering a fixed step size.
Remote command:
FREQ:CENT:STEP 120MHz
Softkeys in time domain:
0.1 * RBW
The
0.1 * RBW
softkey sets the step size for the center frequency entry to 10% of
the resolution bandwidth.
AUTO 0.1 * RBW
corresponds to the default setting.
Remote command:
FREQ:CENT:STEP:LINK RBW
FREQ:CENT:STEP:LINK:FACT 10PCT
0.5 * RBW
The
0.5 * RBW
softkey sets the step size for the center frequency entry to 50% of
the resolution bandwidth.
Remote command:
FREQ:CENT:STEP:LINK RBW
FREQ:CENT:STEP:LINK:FACT 50PCT
X * RBW
The
X * RBW
softkey allows the factor defining the center frequency step size to be
entered as % of the resolution bandwidth.
Values between 1 and 100% in steps of 1% are allowed. The default setting is 10%.
Remote command:
FREQ:CENT:STEP:LINK RBW
FREQ:CENT:STEP:LINK:FACT 20PCT
= CENTER
The
= CENTER
softkey sets the step size coupling to MANUAL and the step size to
the value of the center frequency. This function is especially useful during measure-
ments of the signal harmonic content because by entering the center frequency
each stroke of the
STEP
key selects the center frequency of another harmonic.
Remote command:
--
= MARKER
The
= MARKER
softkey sets the step size coupling to MANUAL and the step size to
the value of the marker. This function is especially useful during measurements of
the signal harmonic content at the marker position because by entering the center
frequency each stroke of the
STEP
key selects the center frequency of another har-
monic.
Remote command:
--
MANUAL
The
MANUAL
softkey activates the window for entering a fixed step size.
Remote command:
FREQ:CENT:STEP 120MHz
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...