CAL CORR
(ON OFF)
The
CAL CORR (ON OFF)
softkey switches the calibration data on/off.
• ON: The status message depends upon the results of the total calibration.
• OFF: The message
UNCAL
appears in the status line.
Remote command:
CAL:STAT ON
YIG CORR
(ON OFF)
The
YIG CORR (ON OFF)
softkey switches on or off the automatic, cyclic correction
of the temperature-dependent frequency drift of the YIG filter.
When switched to ON (default setting), it is checked once per minute whether a fre-
quency correction for the YIG filter is required. Frequency correction is performed if
the temperature has changed by more than 3K relative to the last instance of correc-
tion.
Remote command:
CORR:YIG:TEMP:AUTO ON | OFF
CAL RESULTS
The
CAL RESULTS
softkey calls the
CALIBRATION RESULTS
table, which shows
the correction data found during calibration.
The
CALIBRATION RESULTS
table contains the following information:
– date and time of last record of correction values
– overall results of correction value record
– list of found correction values according to function/module
The results have the following meaning:
The
YIG CORR (ON OFF)
softkey is available only on models with a frequency
range
≥
3 GHz and
≤
40 GHz.
If the instrument is operated in a temperature-controlled environment, the cyclic
frequency drift correction can – for time-critical applications – be switched off after
an operating period of
≥
30 minutes.
PASSED
calibration successful without any restrictions
CHECK
deviation of correction value larger than expected, correction
could however be performed
FAILED
deviations of correction value too large, no correction was possi-
ble. The found correction data are not valid.
ABORTED
calibration aborted
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...