R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
6.24
Operating Manual 1164.4556.12 - 06
6.5.4
CALCulate:FLINe Subsystem
The CALCulate:FLINe subsystem controls the settings for the frequency limit lines. The measurement
windows are selected by CALCulate 1 (screen A) or 2 (screen B).
CALCulate<1|2>:FLINe<1|2>
0 to f
max
This command defines the position of the frequency lines.
The frequency lines mark the frequencies specified in the measurement window. Frequency lines are
only available with SPAN > 0.
Example:
"CALC:FLIN2 120MHz"
Characteristics:
*RST value: - (STATe to OFF)
SCPI: device-specific
Mode:
A-F
CALCulate<1|2>:FLINe<1|2>:STATe
ON | OFF
This command switches the frequency line on or off.
Example:
"CALC:FLIN2:STAT ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A-F
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...