![Rohde & Schwarz R&S FSP Series Скачать руководство пользователя страница 69](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864069.webp)
Operating Manual 1164.4556.12 - 06
2.45
R&S FSP
Getting Started
Measurements on Modulated Signals
➢
Set the
APD
softkey to
ON
.
The R&S FSP sets the frequency span to 0 Hz and measures the amplitude
probability distribution (APD). The number of uncorrelated level
measurements used for the measurement is 100000. The mean power and the
peak power are displayed in dBm. The crest factor (peak power – mean power)
is output as well (see
Fig. 2-1
Amplitude probability distribution of white noise
3. Switch to the CCDF display mode.
➢
Set the
CCDF
softkey to
ON
The APD measurement is switched off and the CCDF display mode is switched
on.
Fig. 2-2
The CCDF of white noise
The CCDF trace indicates the probability that a level will exceed the mean
power. The level above the mean power is plotted along the X axis of the
graph.The origin of the axis corresponds to the mean power level. The
probability that a level will be exceeded is plotted along the Y axis.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...