Operating Manual 1164.4556.12 - 06
2.43
R&S FSP
Getting Started
Measurements on Modulated Signals
Optimum Level Setting for ACP Measurements on W-CDMA Signals
The dynamic range for ACPR measurements is limited by the thermal noise floor,
the phase noise and the intermodulation (spectral regrowth) of the R&S FSP. The
power values produced by the R&S FSP due to these factors accumulate linearly.
They depend on the applied level at the input mixer. The three factors are shown in
the figure below for the adjacent channel (5 MHz carrier offset)
Fig. 2.20 The R&S FSP’s dynamic range for adjacent channel power measurements on W-CDMA
uplink signals is a function of the mixer level.
The level of the W-CDMA signal at the input mixer is shown on the horizontal axis,
i.e. the measured signal level minus the selected RF attenuation. The individual
components which contribute to the power in the adjacent channel and the resulting
relative level (total ACPR) in the adjacent channel are displayed on the vertical axis.
The optimum mixer level is –21 dBm. The relative adjacent channel power (ACPR)
at an optimum mixer level is –65 dBc. Since, at a given signal level, the mixer level
is set in 10 dB steps with the 10 dB RF attenuator, the optimum 10 dB range is
shown in the figure: it spreads from -16 dBm to -26 dBm. The obtainable dynamic
range in this range is 62 dB.
To set the attenuation parameter manually, the following method is recommended:
• Set the RF attenuation so that the mixer level (= measured channel power – RF
attenuation) is between -16 dBm and -26 dBm.
• Set the reference level to the largest possible value where no overload (IFOVLD)
is indicated.
This method is automated with the R&S FSP’s ADJUST REF LEVEL function. Espe-
cially in remote control mode, e.g. in production environments, it is best to correctly
set the attenuation parameters prior to the measurement, as the time required for
automatic setting can be saved.
-80
-75
-70
-65
-60
-55
-50
-45
-40
-35
-30
-40
-35
-30
-25
-20
-15
-10
Mixer Level / dBm
ACLR / dBc
Total ACLR
Phase Noise
Thermal Noise
S.R.I.
optimum 10-dB range
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...