R&S FSP
Remote Control – Description of Commands
SYSTem Subsystem
6.230
Operating Manual 1164.4556.12 - 06
SYSTem:LANGuage
<language>
This command activates the emulation of various analyzers. The analyzer default command set is
SCPI.
The following remote-control languages are available:
SCPI, 71100C, 71200C, 71209A, 8560E, 8561E, 8562E, 8563E, 8564E, 8565E, 8566A, 8566B,
8568A, 8568A_DC, 8568B, 8568B_DC, 8591E, 8594E
The settings or changes made on switching between remote-control languages are described in the
Quick Start Guide, chapter 2.
Example:
"SYST:LANG 'SCPI'"
Characteristics:
*RST value: 'SCPI'
SCPI: conform
Mode:
A
There is no query for this command.
SYSTem:LXI:DISPlay
ON|OFF
This command shows or hides the
LXI Observer
dialog box. To use this command, the LXI Class C
functionality must be installed and enabled (for details see
“LXI Class C Functionality” on page 4.316
Example:
"SYST:LXI:DISP ON"
'Shows the
LXI Observer
dialog box.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
• When SCPI is selected, the 8566B/8568B/8594E command set is available in
addition.
• When 8566A, 8566B, 8568A or 8568B is selected, both command sets A and B
are available, as far as they are supported.
• The HP model 8591E is compatible to HP model 8594E,
the HP models 71100C, 71200C, and 71209A are compatible to HP models
8566A/B.
Notes regarding switch over to 8566A/B and 8568A/B:
• Commands IP and KST, too, will perform the settings for the "# of Trace Points",
"Start Freq.", "Stop Freq." and "Input Coupling".
• Switch over of the "# of Trace Points" will not take place until the instrument is
switched to the remote control mode. For manual operation (selected with
LOCAL softkey), the number of sweep points (trace points) will always be set to
1251.
• In the remote control mode, the screen area for the measurement will be
reduced. The LOCAL softkey (lowermost softkey) will be shifted slightly towards
the center of the screen.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...