Operating Manual 1164.4556.12 - 06
6.175
R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]LIST:POWer:SET:AVERage:TYPE
LINear | LOGarithmic
With this command the average type of the sense list measurement can be set either to linear (default)
or logarithmic. In linear mode voltages are averaged in logarithmic mode levels.
Example:
"LIST:POW:SET:AVER LOG"
'sets log averaging for list measurement.
Characteristics:
*RST value: LIN
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:POWer:STATe
OFF
This command deactivates the list measurement.
Example:
"SENSe:LIST:POWer:STATe OFF"
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:BANDwidth[:RESolution]
<numeric_value>
This command selects the resolution bandwidth (RBW) of a range in the spurious measurement.
Example:
"LIST:RANG2:BAND 10E3"
'Sets the RBW to 10 KHz.
Characteristics:
*RST value: 10 kHz
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:BANDwidth:VIDeo
<numeric_value>
This command selects the video bandwidth (VBW) of a range in the spurious measurement.
Example:
"LIST:RANG2:BAND:VIDeo 40E3"
'Sets the VBW to 40 KHz.
Characteristics:
*RST value: 30 kHz
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:BREak
ON | OFF
This command determines whether a sweep is stopped in the spurious measurement when a range is
changed.
Example:
"LIST:RANG2:BRE ON"
'Sweep stops if the range is changed from range 2 to 3.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...