R&S FSP
Remote Control – Description of Commands
SENSe Subsystem
6.158
Operating Manual 1164.4556.12 - 06
6.16.4
SENSe:DETector Subsystem
The SENSe:DETector subsystem controls the acquisition of measurement data via the selection of the
detector for the corresponding trace. The measurement windows are selected by SENSe1 (screen A)
and SENSe2 (screen B).
[SENSe<1|2>:]DETector<1..3>[:FUNCtion]
APEak | NEGative | POSitive | SAMPle | RMS | AVERage
| QPEak
This command switches on the detector for the data acquisition in the selected trace and the indicated
measurement window.
• The APEak detector (AutoPeak) displays the positive and also the negative peak value of the noise
floor. If a signal is detected, only the positive peak value is displayed.
• The POSitive or NEGative detector only displays the positive or the negative peak value.
• With the Sample detector the value measured at the sampling time is displayed, whereas the RMS
value of the power measured at each measurement point is displayed with the RMS detector.
• The AVERage detector displays the power average value at each measurement point.
• The QPEak detector performs a signal evaluation for EMC measurements.
If QPEak is selected, the video filter is automatically switched off. The couplings between span and
RBW as well as between RBW and sweep time are also switched off and restored on selecting another
detector. A long sweep time should be selected so that the quasipeak detector can fully settle at each
measurement point.
The trace is indicated as numeric suffix in
DETector
.
Example:
"DET POS"
'Sets the detector in screen A to "positive peak".
Characteristics:
*RST value: APEak
SCPI: conform
Mode:
A
[SENSe<1|2>:]DETector<1..3>[:FUNCtion]:AUTO
ON | OFF
This command either couples the detector in the selected measurement window to the current trace
setting or turns coupling off. The trace is selected by the numeric suffix at DETector.
Example:
"DET:AUTO OFF"
Characteristics:
*RST value: ON
SCPI: conform
Mode:
A
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...