Operating Manual 1164.4556.12 - 06
4.285
R&S FSP
Instrument Functions
LO/IF ports for external mixers - Option R&S FSP-B21
4. Level correction
The conversion loss of the mixer can be taken into account both as a function of
frequency and with an average value. The frequency-dependent level correction
is used in this example for obtaining a higher precision. First select a table valid
for the selected band. As an alternative, level correction based on average value
can be used (see
“Level correction with average value” on page 4.286
5. Frequency-dependent level correction
➢
Activate a selection list which contains all tables stored on the hard disk of the
R&S FSP with
[
FREQUENCY
: EXTERNAL MIXER : SELECT BAND]
➢
Select the
CONV LOSS TABLE
field by means of the cursor keys.
➢
If a file with correction data already exists for the mixer used, select a table
valid with the cursor keys or the rotary knob and confirm with
ENTER
.
➢
If such a table is not yet available, create a new table with
[CONV LOSS TABLE : EDIT TABLE]
or copy from a disk to the hard disk of the R&S FSP with
[CONV LOSS TABLE : LOAD TABLE]
(See section
“Conversion Loss Tables” on page 4.271
and select the file created or loaded from a disk in the selection list as
described before.
A selected file is marked by
√
and will be drawn into the table SELECT BAND.
No further settings are necessary since the selected file contains all required
parameters.
➢
Quit the
SELECT BAND
menu using
[
ESC
] key
A span is automatically set which covers the whole V band (50 to 75 GHz).
➢
Set the frequency range to be examined using
[ FREQUENCY START : 52 GHz ]
and
[ FREQUENCY STOP : 60 GHz ]
➢
Reduce video bandwidth with
[
SWEEP COUPLING
: VIDEO BW MANUAL :
1 MHz
]
A correct signal identification using
AUTO ID
is thus possible (see also
“Remarks Concerning Signal Identification with AUTO ID” on page 4.278
The display shown in
is obtained.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...