![Rohde & Schwarz R&S FSP Series Скачать руководство пользователя страница 615](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864615.webp)
Operating Manual 1164.4556.12 - 06
6.221
R&S FSP
Remote Control – Description of Commands
SYSTem Subsystem
6.19 SYSTem Subsystem
This subsystem contains a series of commands for general functions.
SYSTem:APPLication:SRECovery[:STATe]
ON | OFF
This command controls the instrument behaviour when changing the active application, e.g from
SPECTRUM to FM DEMOD and back from FM DEMOD to SPECTRUM. In the default state OFF a
few parameters of the current analyzer setting are passed to the application (e.g. center frequency,
level settings) or from the application back to the analyzer mode. If APPL SETUP RECOVERY is
switched ON, the settings of the applications are independent of each other. Leaving the FM DEMOD
application will restore the previous state of the ANALYZER.
Example:
"SYST:APPL:SREC ON
'use independent settings
Characteristics:
*RST value: OFF
SCPI:device-specific
Mode:
A
SYSTem:COMMunicate:GPIB:RDEVice:COMMand
0...30,<’command string’>
This command sends a command or query string to the external generator connected via the GPIB
interface of the Ext. Generator Control, R&S FSP-B10. Usually an external generator is connected to
this interface.
Parameter:
0...30: GPIB address of the external generator connected via the GPIB interface of
the Ext. Generator Control, R&S FSP-B10. In the following example, the GPIB
address is 18.
Example:
"SYST:COMM:GPIB:RDEF:COMM 18,'*RST'"
’Resets generator.
"SYST:COMM:GPIB:RDEF:COMM 18,'SOURCE:FREQ:CW 2E9'"
’Sets generator frequency to 2 GHz.
"SYST:COMM:GPIB:RDEF:COMM 18,'SOUR:POW 0'"
'Sets generator power to 0dBm.
"SYST:COMM:GPIB:RDEF:COMM? 18,'SENS:POW?'"
'Queries the generator power.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
The individual application settings are stored on the internal harddisk.
Only basic I/O is possible.
The first parameter is the GPIB address of the device connected with GPIB inter-
face of option FSP-B10.
The command is only available with option Ext. Generator Control, R&S FSP-B10.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...