Operating Manual 1164.4556.12 - 06
2.35
R&S FSP
Getting Started
Measurements on Modulated Signals
5. Setting the optimum reference level and RF attenuation for the
measurement
➢
Press the
ADJUST REF LEVEL
softkey.
The R&S FSP sets the optimum RF attenuation and the optimum reference
level on the basis of the measured channel power.
Fig. 2-1
Measuring the relative adjacent channel power of an NADC signal in each
of the two adjacent channels below and above the transmit channel.
To increase repeatability – especially in the adjacent channels – the R&S FSP’s Fast
ACP routine is recommended.
6. Switching on the Fast ACP routine.
➢
Press the
CP/ACP CONFIG
!
softkey
➢
Set the
FAST ACP
softkey to
ON.
➢
Press the
ADJUST REF LEVEL
softkey.
The R&S FSP makes consecutive measurements on the 5 channels in the
Zero Span mode using the receive filter specified in IS 136 to define the
resolution bandwidth. The power in each channel is displayed on the graph as
a function of time
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...