Operating Manual 1164.4556.12 - 06
6.111
R&S FSP
Remote Control – Description of Commands
DISPlay Subsystem
Example:
"INIT:CONT OFF"
'Switching to single-sweep mode.
"SWE:COUN 16"
'Sets the number of measurements to 16.
"DISP:WIND1:TRAC3:MODE MAXH"
'Switches on the calculation of the for trace 3 in screen A.
"INIT;*WAI"
'Starts the measurement and waits for the end of the 16 sweeps.
Characteristics:
*RST value: WRITe for TRACe1, STATe OFF for TRACe2/3
SCPI: device-specific
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:MODE:HCONtinuous
ON | OFF
This command specifies whether or not the traces with peak or minimum value detection are reset after
specific parameter changes.
Usually the measurement must be restarted after a parameter change, before an evaluation of the
measurement results is performed (e.g. with a marker). In cases in which a change causes a
compulsory new measurement, the trace is automatically reset in order to prevent erroneous
measurements of previous measurement results (e.g. when the span changes). For applications in
which this behavior is not desired, this mechanism can be switched off.
Parameter:
OFF: The traces are reset after specific parameter changes.
ON: The reset mechanism is switched off.
Example:
"DISP:WIND1:TRAC3:MODE:HCON ON"
'The reset mechanism is switched off for measurement window 1.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>[:STATe]
ON | OFF
This command switches on or off the display of the corresponding trace in the selected measurement
window.
Example:
"DISP:WIND1:TRAC3 ON"
Characteristics:
*RST value: ON for TRACe1, OFF for TRACe2 and 3
SCPI: conform
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:X[:SCALe]:ZOOM
ON | OFF
This command switches the zoom on or off.
Example:
"DISP:TRAC:X:ZOOM ON"
Characteristics:
*RST value: OFF
SCPI: conform
Mode:
A
The numeric suffix in TRACe<1...3> is irrelevant.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...