Operating Manual 1164.4556.12 - 06
4.243
R&S FSP
Instrument Functions
External Generator Control – Option R&S FSP-B10
– In the time domain (Span = 0 Hz) the generator frequency is derived from the
set receive frequency of the R&S FSP using the calculation formula.
For the sake of clarity, the formula is also displayed in the table.
The offset setting can be used to sweep in the reverse direction. This can be
achieved by setting a
negative
offset in the formula above:
Example for reverse sweep:
F
AnalyzerStart
= 100 MHz
F
AnalyzerStop
= 200 MHz
F
Offset
= -300 MHz
Numerator
=
Denominator
= 1
→
F
GeneratorStart
= 200 MHz
→
F
GeneratorStop
= 100 MHz
If the offset is adjusted in a way that the sweep of the generator crosses the 0 Hz
Frequency, it is indicated by the additional statement “via 0 Hz”.
Example for reverse sweep via 0 Hz
F
AnalyzerStart
= 100 MHz
F
AnalyzerStop
= 200 MHz
F
Offset
= -150 MHz
Numerator
=
Denominator
= 1
→
F
GeneratorStart
= 50 MHz
→
F
GeneratorStop
= 50 MHz via 0 Hz
Remote command:
:
SOUR:EXT:FREQ:NUM 1
:
SOUR:EXT:FREQ:DEN 1
:
SOUR:EXT:FREQ:OFFS -300MHZ
• RESULT
The frequency range of the generator resulting from the calculation formula. An
asterisk (*) after the upper limit indicates that the stop frequency of the R&S FSP
must be adapted when the generator is switched on in order not to exceed its
maximum frequency. In the following illustration, this is true for the upper
generator at a stop frequency of 3.2 GHz of the R&S FSP, whereas the lower
generator does not yet require an adaptation:
Remote command:
SOUR:EXT:POW –30dBm
SOUR:EXT:FREQ:NUM 4
SOUR:EXT:FREQ:DEN 3
SOUR:EXT:FREQ:OFFS 100MHZ
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...