R&S FSP
Instrument Functions
External Generator Control – Option R&S FSP-B10
4.238
Operating Manual 1164.4556.12 - 06
EXT SRC ON /
OFF
The
EXT SRC ON / OFF
softkey switches the external generator on or off.
It can only be switched on successfully if the generator has been selected by means
of
SELECT GENERATOR
and configured correctly by means of
FREQUENCY
SWEEP
. If one of these conditions is not fulfilled, an error message will be output.
Remote command:
SOUR:EXT ON
SELECT
GENERATOR
The
SELECT GENERATOR
softkey opens a table for selection of the generator and
definition of GPIB address and control interface.
The table permits configuration of two generators so that switching between two dif-
ferent configurations is easily possible.
The individual fields contain the following settings:
• SRC
Index of generator selected
When switching on the external generator by means of
EXT SRC ON
, the
R&S FSP switches off the internal tracking generator and starts programming the
generator settings via the GPIB interface IEC2.
Programming requires takeover of the remote control at this interface by the
R&S FSP. To avoid any access conflicts, ensure that no other controller is con-
nected to the IEC2 interface or the external generator when selecting
EXT SRC
ON
.
The maximum stop frequency is limited to the maximum generator frequency. This
upper limit is automatically reduced by the set frequency offset of the generator
and a set multiplication factor.
With the external generator switched on, the FFT filters (
FILTER TYPE FFT
in the
menu
BW
) are not available.
If an error occurs on the GPIB when programming the external generator, the gen-
erator will automatically be switched off and the following error message will be
output:
When the external generator is switched off using
EXT SRC OFF
, the GPIB control
is handed over again at the IEC2 interface, i.e. a different controller will then take
over the control of the signal generator.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...