R&S FSP
Instrument Functions
Analyzer Mode
4.58
Operating Manual 1164.4556.12 - 06
STEPSIZE
STANDARD
The
STEPSIZE STANDARD
softkey controls the knob increment/decrement of the
marker position and uses the grid resolution (span/625).
Remote command:
CALC:MARK:X:SSIZ STAN
STEPSIZE
SWP POINTS
The
STEPSIZE SWP POINTS
softkeys controls the knob increment/decrement of
the marker position and uses the available sweep points configured in the
SWEEP
menu.
Remote command:
CALC:MARK:X:SSIZ POIN
MKR FILE
EXPORT
The
MKR FILE EXPORT
softkey saves the data of all active markers of the window
to a specified file. The format of the decimal point is defined by the
DECIM SEP
soft-
key.
Remote command:
MMEM:STOR:MARK 'C:\marker.txt'
DECIM SEP
The
DECIM SEP
softkey selects the decimal separator between '.' (decimal point)
and ',' (comma) for the function
MKR FILE EXPORT
.
Remote command:
FORM:DEXP:DSEP POIN
4.5.11
Marker Functions – MKR FCTN Key
The
MKR FCTN
menu offers further measurements with the markers:
– Measurement of noise density (
softkey)
– Measurement of phase noise (
softkey)
– Measurement of filter or signal bandwidth (
softkey)
– Activating of AF demodulation (
softkey)
On calling the menu, the entry for the last active marker is activated (
softkey); if no marker is activated, marker 1 is activated and a maximum
search (
softkey) is performed. The marker can be set to the desired trace by
means of
softkey.
[
SIGNAL COUNT
]
The frequency counter is switched on. R&S FSP counts the
frequency of the signal at the marker position with a resolu-
tion of 1 kHz. The counted frequency is indicated in the
marker info field.
[
NEXT
]
Changes to the submenu for setting the counter resolution.
[
CNT RESOL 10 Hz
]
The frequency counter resolution is increased to 10 Hz.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...