Operating Manual 1164.4556.12 - 06
4.55
R&S FSP
Instrument Functions
Analyzer Mode
Additionally, the
REFERENCE FIXED
softkey opens the submenu where it is possi-
ble to determine manually a reference point with level and frequency, time or x-axis
level, to define a level offset or deactivate the reference point.
The
REFERENCE FIXED
function is useful for the measurement of the harmonic
suppression at small span (fundamental not represented).
REF FXD
(ON OFF)
The
REF FXD (ON OFF)
softkey switches on or off the relative measurement to a
fixed reference value (
REFERENCE POINT
) independent of the trace.
Remote command:
CALC:DELT2:FUNC:FIX ON
REF POINT
LEVEL
The
REF POINT LEVEL
softkey enters a reference level independent of the refer-
ence marker level. All relative level values of the delta markers refer to this refer-
ence level.
Remote command:
CALC:DELT2:FUNC:FIX:RPO:Y -10dBm
REF POINT LVL
OFFSET
The
REF POINT LVL OFFSET
softkey specifies a level offset relevant to the refer-
ence level. The relative level values of the delta markers refer to the reference point
level plus the level offset.
The level offset is set to 0 dB on enabling the
REFERENCE FIXED
or
PHASE
NOISE
function.
Remote command:
CALC:DELT2:FUNC:FIX:RPO:Y:OFFS 0dB
REF POINT
FREQUENCY
With the
REF POINT FREQUENCY
softkey a reference frequency can be manually
activated for the delta markers when the
REFERENCE FIXED
or
PHASE NOISE
function is used.
Remote command:
CALC:DELT2:FUNC:FIX:RPO:X 10.7MHz
REF POINT
TIME
The
REF POINT TIME
softkey activates the entry box for the input of a reference
time for the
REFERENCE FIXED
function in the time domain (span = 0 Hz).
Remote command:
CALC:DELT2:FUNC:FIX:RPO:X 5MS
For phase noise measurement, input of reference time is not possible.
PEAK SEARCH
The
PEAK SEARCH
softkey defines the maximum of the selected trace as refer-
ence level for all delta markers when using the
REFERENCE FIXED
function.
Remote command:
CALC:DELT2:FUNC:FIX:RPO:X -5DBM
Measurement example
Small-span harmonics measurement to increase sensitivity
CW signal (e.g. 100 MHz, 0 dBm) with harmonics at the RF input of R&S FSP.
[
PRESET
]
R&S FSP is set to the default setting.
[
CENTER
: 100
MHz
]
The center frequency of R&S FSP is set to 100 MHz.
[
SPAN
: 1
MHz
]
The span is set to 1 MHz.
[
AMPL
: 3
dBm
]
The reference level is set to 3 dBm (3 dB above the
expected RF level).
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...