![Rohde & Schwarz R&S FSP Series Скачать руководство пользователя страница 248](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864248.webp)
R&S FSP
Instrument Functions
Basic Settings
4.176
Operating Manual 1164.4556.12 - 06
Remote command:
SYST:LANG "SCPI" | "8560E" |
"8561E" | "8562E" | "8563E" |
"8564E" | "8565E" | "8566A" |
"8566B" | "8568A" | "8568A_DC" |
"8568B" | "8568B_DC" | "8591E" |
"8594E" | "71100C" | "71200C" |
"71209A"
IF GAIN
(NORM PULS)
The
IF GAIN (NORM PULS)
softkey configures the internal IF gain settings in HP
emulation mode due to the application needs. This setting is only taken into account
for resolution bandwidth < 300 kHz.
This softkey is only available, if a HP languge is selected via the
softkey.
Remote command:
SYST:IFG:MODE PULS
SWEEP REP
(ON OFF)
The
SWEEP REP (ON OFF)
softkey controls a repeated sweep of the
E1
and
MKPK
HI
HP model commands (for details on the commands refer to
HP Models 856xE, 8566A/B, 8568A/B and 8594E” on page 6.253
). If the repeated
sweep is OFF, the marker is set without sweeping before.
This softkey is only available, if a HP languge is selected via the
softkey.
Remote command:
SYST:RSW ON
COUPLING
(FSP HP)
The
COUPLING (FSP HP)
softkey controls the default coupling ratios for:
• span and resolution bandwidth (Span/RBW) and
• resolution bandwidth and video bandwidth (RBW/VBW)
for the HP emulation mode.
In case of FSP the standard parameter coupling of the instrument is used. As a
result in most cases a shorter sweeptime is used than in case of HP.
The softkey is only available in HP emulation mode.
Remote command:
SYST:HPC FSP
NORM
Optimized for high dynamic range,
overload limit is close to reference level.
PULS
Optimized for pulsed signals,
overload limit up to 10 dB above reference level.
In single sweep mode, switch off this softkey before you set the marker via the
E1
and
MKPK HI
commands in order to avoid sweeping afresh.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...