![Rohde & Schwarz R&S FSP Series Скачать руководство пользователя страница 413](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864413.webp)
Operating Manual 1164.4556.12 - 06
6.19
R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:DELTamarker<1...4>:MINimum:RIGHt
This command positions the specified delta marker to the next higher minimum value to the right of the
current value (i.e. ascending X values). The corresponding delta marker will be activated first, if
necessary.
Example:
"CALC2:DELT:MIN:RIGH"
'Sets delta marker 1 in screen B to the next higher minimum value to the right of
the current value.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1...4>:MODE
ABSolute | RELative
This command switches between relative and absolute frequency input of the delta marker (or time
with span = 0). It affects all delta markers independent of the measurement window.
Example:
"CALC:DELT:MODE ABS"
'Switches the frequency/time indication for all delta markers to absolute values.
"CALC:DELT:MODE REL"
'Switches the frequency/time indication for all delta markers to relative to marker 1.
Characteristics:
*RST value: REL
SCPI: device-specific
Mode:
A
Mode:
CALCulate<1|2>:DELTamarker<1...4>[:STATe]
ON | OFF
This command switches on and off the delta marker if delta marker 1 is selected. If marker 2, 3 or 4 is
selected and used as a marker, it is switched to delta marker mode. If the corresponding marker is not
activated, it will be activated and positioned on the maximum of the measurement curve.
If no numeric suffix is indicated, delta marker 1 is selected automatically.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Example:
"CALC:DELT3 ON"
'Switches marker 3 in screen A to delta marker mode.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...