timeout has occurred. For more information, see Section 6.3, “Automatic alerts and
notification of clinical management and system performance events”, page 133.
Balancing formation interval with longevity – A shorter Minimum Auto Cap Formation
Interval forms the capacitors more frequently, which optimizes the efficiency of the
capacitors. However, each capacitor formation is produced by a full energy charge, which
reduces the longevity of the device. Assess the requirements of a patient for faster therapy
delivery relative to the effect on device longevity. For information about device longevity,
see Section 6.15, “Optimizing device longevity”, page 197.
9.9.4 Programming the automatic capacitor formation interval
Select Params icon
⇒ VF | Therapies…
⇒ Auto Cap Formation…
▷ Minimum Auto Cap Formation Interval
9.9.5 Evaluation of charge time
To evaluate the current charge performance of the device, review the charge time and date
displayed on the Battery and Lead Measurements screen. For more information, see
Section 6.13, “Viewing detailed device and lead performance data”, page 186. If the Last
Charge date is recent and the charge time is acceptable, there is no need for further
investigation.
Figure 219. Charge time data on the Battery and Lead Measurements screen
Otherwise, use the Charge/Dump Test to assess the charge time. For more information,
see Section 10.6, “Testing the device capacitors”, page 434.