8.4.5 Evaluation of Wavelet
8.4.5.1 Wavelet Test
You can use the Wavelet Test to evaluate the accuracy of the current Wavelet template and,
if necessary, to manually collect an updated template.
For a full description of the Wavelet Test, see Section 10.3, “Testing the Wavelet feature”,
8.4.5.2 Wavelet Monitor
You can use the Wavelet Monitor mode to evaluate the potential effectiveness of Wavelet
for the patient without enabling it for detection. When Wavelet has been programmed to
Monitor, the device records Wavelet-related data but does not use the Wavelet feature to
withhold detection.
8.4.5.3 Data - Arrhythmia Episodes screen
The stored episode data provides information related to the operation of Wavelet.
Select Data icon
⇒ Clinical Diagnostics
⇒ Arrhythmia Episodes
Medtronic