Figure 159. VT/VF Episode counter
8.3 Discriminating VT/VF from SVT using PR Logic
Patients who are experiencing supraventricular tachycardia (SVT) may exhibit ventricular
rates in the VT/VF detection zone. If sustained, such fast ventricular rates may cause an
inappropriate delivery of a tachyarrhythmia therapy. Identifying and withholding detection
for conducted SVT reduces the chance of delivering an inappropriate therapy for high
ventricular rates that are not ventricular in origin.
8.3.1 System solution: PR Logic
PR Logic uses pattern and rate analysis to discriminate between SVTs and true ventricular
tachyarrhythmias and to withhold inappropriate VT/VF detection and therapy during
episodes of rapidly conducted SVT.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
321