6.13.4 Viewing capture threshold trends
If Capture Management is programmed to Adaptive or Monitor, the device automatically
performs daily pacing threshold searches and records the results in the capture threshold
trends data. For more information about Capture Management, see Section 7.5, “Managing
pacing output energies with Capture Management”, page 238.
The results of the daily pacing threshold measurements are displayed on the Lead Trends
screen in the Capture Threshold trend graph. The graph displays up to 15 of the most recent
measurements and up to 80 weekly summary measurements (showing minimum,
maximum, and average values for each week).
The Lead Trends screen also displays programmed values for pacing output and Capture
Management parameters, the last measured threshold value, and a link to a detailed view
of the last 15 days of threshold measurement data. The details screen shows daily results
from the last 15 days of threshold measurements. These results include the dates, times,
threshold measurements, pacing amplitude values, and notes describing the results of each
pacing threshold search.
The capture threshold trend data provides a way to evaluate the operation of Capture
Management and the appropriateness of the current pacing output values. In addition,
sudden or significant changes in pacing threshold may indicate a problem with a lead.
6.13.4.1 How to view capture threshold trends
Select Data icon
⇒ Device/Lead Diagnostics
⇒ Capture Threshold Trends