![Medtronic PROTECTA XT DR D314DRG Скачать руководство пользователя страница 379](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148379.webp)
VT/VF Therapy Summary – This counter reports the number of pace-terminated
tachyarrhythmias, shock-terminated tachyarrhythmias, total VT/VF shocks, and aborted
charges for the prior session, the last session, and the device lifetime.
VT/VF Therapy Efficacy Since Last Session – This counter reports the number and types
of VF, FVT, and VT therapies that were delivered and successful. The VT Therapy counter
includes VT episodes that accelerated during the therapy or were redetected as an FVT or
VF episode. The FVT therapy counter includes FVT episodes that were redetected as a VF
episode.
9.2.5.3 Smart Mode operation indicators
The VT/VF therapy counters screen provides information about the operation of Smart
Mode.
Select Data icon
⇒ Clinical Diagnostics
⇒ Counters
⇒ VT/VF Rx
Figure 194. VT/VF therapy counters screen
1 The label “Off-SM” in the VT Therapy counter indicates that Smart Mode cancelled an
unsuccessful ATP therapy.
Information about Smart Mode operation is also available from the VT Therapies screen or
the FVT Therapies screen.