EGM data storage and device memory conservation – For VT/VF or VT Monitor
episodes, the device begins to store ventricular EGM data when 3 consecutive intervals
have occurred in the VT, VT Monitor, or VF zone.
For AT/AF episodes, the device begins to store atrial EGM data when the device detects
AT/AF Onset. The device stores up to 5 s of EGM data prior to AT/AF detection, regardless
of whether a Pre-arrhythmia EGM storage option is selected. For more information about
Pre-arrhythmia EGM storage, see Section 6.8.4, “How to set data collection preferences”,
page 173.
To conserve device memory, the EGM is stored only during specific parts of an episode.
Note: Long episodes may contain gaps in the EGM to save storage memory.
6.8.3.3 Viewing the episode text
When you select an episode from the episode log and then select the Text option, the
programmer displays a text summary of the episode.
Figure 49. Episode Text
1 Use the vertical scroll bar on the right side of the screen to scroll through all of the episode text.
Medtronic
PROTECTA™ XT DR D314DRG
172
Clinician Manual