10.7.4.1 How to deliver an atrial 50 Hz Burst
Figure 222. Atrial 50 Hz Burst induction screen
1. Select Tests > EP Study.
2. Select 50 Hz Burst from the list of inductions and therapies.
3. If the Select Chamber dialog box appears, select [Atrium]. Otherwise, make sure that
the Chamber parameter is set to Atrium.
4. If you want to treat the induced episode with a manual therapy, select [Suspend] to
prevent automatic detection.
Note: The [Suspend] button is located at the top of the screen and is not shown in
5. Select the Resume at BURST check box for automatic detection and therapy, or clear
the check box for manual therapy.
6. Accept the displayed test values or select new test values.
7. If you want to provide VOO Backup pacing during the pacing burst, select values for
VOO Backup.
8. Press and hold [50 Hz BURST Press and Hold]. Release the button to end the
induction.
9. If necessary, select [ABORT] to abort a therapy in progress.
10.7.5 Inducing AT or VT with Fixed Burst
You may use the Fixed Burst inductions to induce AT or VT. To induce atrial or ventricular
tachyarrhythmias, the Fixed Burst induction delivers a set of asynchronous AOO or VOO
pacing pulses at a uniform, selectable interval to the designated chamber. You may also
specify the amplitude and pulse width of the pulses.