Figure 117. Evaluating Mode Switch operation during an AT/AF episode
1 The low average ventricular rate and the difference between the average atrial and ventricular
rates suggest that the device is not operating in a tracking mode.
2 The stored EGM shows that the device is not tracking the atrial rate when pacing the ventricle.
7.15.5.2 Mode Switch transitions
The Marker Channel includes an “MS” marker for each Mode Switch transition, either to a
nontracking mode or back to a tracking mode.
The current operating mode is displayed in the upper left-hand corner of the screen. During
a Mode Switch episode, DDIR is displayed.
7.16 Increasing the pacing output after a high-voltage
therapy
After the heart receives a high-voltage therapy, there may be a temporary rise in the pacing
thresholds. The rise in the pacing threshold may result in loss of capture.