4.2.1.1 How to start a patient session using wireless telemetry
1. Select [Find Patient…] from the Select Model window.
2. Select the “Allow wireless communication” check box on the Find Patient window.
3. Use the Conexus Activator, or briefly place the programming head over the device to
activate wireless telemetry in the device.
Notes:
●
When the Conexus Activator is used to activate telemetry in the device, the
programmer launches the patient session without suspending tachyarrhythmia
detection. Placing a magnet near the device, however, suspends tachyarrhythmia
detection.
●
When the programming head is used to activate telemetry in the device, the
programmer automatically launches the patient session with tachyarrhythmia
detection suspended. Detection remains suspended as long as the programming
head is over the device. A warning reminds you that tachyarrhythmia detection is
suspended.
4. Select the appropriate patient from the Patient Name list on the Find Patient window.
Note: The programmer lists all patients with wireless-activated implantable devices
within telemetry range.
5. Select [Start].
4.2.1.2 How to start a patient session using nonwireless telemetry
1. Select [Find Patient…] from the Select Model window.
2. If you are using a Medtronic CareLink Model 2090 Programmer with Conexus
telemetry, make sure that the “Allow wireless communication” check box on the Find
Patient window is not selected. If you start a session with the programming head over
the patient’s device and the “Allow wireless communication” check box is selected, the
system initiates a wireless telemetry session and automatically interrogates the device.
If you are using a Medtronic CareLink Model 2090 Programmer that does not have
Conexus telemetry, the “Allow wireless communication” check box does not appear
on the Find Patient window.
3. Place the programming head over the device and the nonwireless session
automatically begins.