7.3.2.3 AV conduction resumes
After switching to DDDR or DDD mode, the device periodically checks AV conduction for
an opportunity to return to AAIR or AAI mode. The first AV conduction check occurs 1 min
after switching to DDDR or DDD mode. During the conduction check, the device switches
to AAIR or AAI pacing mode for one cycle.
●
If the next A-A interval includes a sensed ventricular beat, the conduction check
succeeds. The device remains in AAIR or AAI pacing mode.
●
If the next A-A interval does not include a sensed ventricular beat, the conduction check
fails, and the device switches back to the DDDR or DDD mode. The time between
conduction checks doubles (2, 4, 8 … min, up to a maximum of 16 hours) with each
failed conduction check.
Figure 78. Switching from DDDR mode to AAIR mode after AV conduction resumes
A
P
A
P
A
P
A
P
A
P
A
P
V
P
A
P
A
P
V
P
V
P
V
S
V
S
V
S
V
S
V
S
V-V intervals
ECG
Marker Channel
200 ms
1 The device operates in DDDR mode.
2 The device performs an AV conduction check. AV conduction is detected.
3 The device operates in AAIR mode.
7.3.2.4 Complete AV block
For patients with complete AV block, the device operates in DDDR or DDD mode
persistently. Every 16 hours, the device checks for AV conduction, which results in a single
dropped ventricular beat.