5.1.2 Setting up the programmer and starting the application
See the programmer reference guide for instructions about how to set up the programmer.
The Model SW009 software should be installed on the programmer. Establish telemetry
with the device and start a patient session.
5.1.3 Considerations for preparing for an implant
Review the following information before implanting the leads or device:
Warning: Do not allow the patient to have contact with grounded electrical equipment that
might produce electrical current leakage during implant. Electrical current leakage may
induce tachyarrhythmias that may result in the patient’s death.
Warning: Keep external defibrillation equipment nearby for immediate use. Potentially
harmful spontaneous or induced tachyarrhythmias may occur during device testing, implant
procedures, and post-implant testing.
Caution: The device is intended for implant in the pectoral region with Medtronic
transvenous defibrillation leads. Implanting the device outside of the pectoral region, or
using an epicardial defibrillation lead instead of an RVcoil (HVB) may adversely affect the
results of the OptiVol fluid measurements. No claims of safety and performance can be
made with regard to other acutely or chronically implanted lead systems that are not
manufactured by Medtronic.
Caution: Lead coils and Active Can electrodes that are in contact during a high-voltage
therapy may cause electrical current to bypass the heart, possibly damaging the device and
leads. While the device is connected to the leads, verify that therapeutic electrodes, stylets,
or guide wires are not touching or connected by any material that may conduct electricity.
Move objects made from conductive materials (for example, an implanted guide wire) well
away from all electrodes before delivering a high-voltage shock.
Caution: Do not implant the device after the “Use by” date on the package label. Battery
longevity may be reduced.
5.1.4 How to prepare the device for implant
Before opening the sterile package, perform the following steps to prepare the device for
implant:
1. Interrogate the device and print an Initial Interrogation Report.
Caution: If the programmer reports that an electrical reset occurred, do not implant
the device. Contact a Medtronic representative.
Medtronic
PROTECTA™ XT DR D314DRG
110
Clinician Manual