Figure 171. Onset Decision Channel annotation indicating detection withheld
1 An episode for which detection is withheld is indicated by “SVT-Onset”.
2 In the Decision Channel of the EGM display, ventricular events on which VT detection was
withheld by the Onset feature are indicated by “Reset: Onset”.
8.5.5.2 Onset Monitor option
Select Parameters icon
⇒ Detection (V.) …
⇒ Onset …
▷ Onset <Monitor>
▷ Percent
You can use the Monitor setting to test the potential effectiveness of the Onset feature for
the patient without programming it to On.
When the Onset feature is set to Monitor, the device performs all of the calculations
associated with the Onset feature but the calculations do not affect the classification of VT
intervals. If the device detects a VT or FVT via VT episode for which detection would have
been withheld had the Onset feature been programmed to On, the episode is noted in the
episode text.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
339