Table 30. Tachyarrhythmia detection parameters (continued)
Parameter
Programmable values
Shipped
Reset
Fast AT/AF Interval
150; 160 … 200 … 250 ms
200 ms
200 ms
On ; OFF
OFF
On
240; 250 … 320 … 400 ms
320 ms
320 ms
VF Initial Beats to
Detect
12/16; 18/24 ; 24/32; 30/40;
45/60; 60/80; 75/100; 90/120;
105/140; 120/160
18/24
18/24
VF Beats to Redetect
6/8; 9/12; 12/16 ; 18/24; 21/28;
24/32; 27/36; 30/40
12/16
12/16
FVT Detection
OFF ; via VF; via VT
OFF
OFF
FVT Interval (Rate)
200; 210 … 240 … 600 ms
—
—
VT Detection
On; OFF
OFF
OFF
280; 290 … 360 … 650 ms
360 ms
400 ms
VT Initial Beats to
Detect
12; 16 … 52; 76; 100
16
16
VT Beats to Redetect
8; 12 … 52
12
12
VT Monitor
Monitor ; Off
Off
Off
VT Monitor Interval
280; 290 … 450 … 650 ms
450 ms
450 ms
Monitored VT Beats to
Detect
16; 20 … 56; 80; 110; 130
20
20
PR Logic/Wavelet
AF/Afl
On ; Off
Off
Off
Sinus Tach
On ; Off
Off
Off
Other 1:1 SVTs
On; Off
Off
Off
Wavelet…
Wavelet
On ; Off; Monitor
Off
Off
Template
[date]
None
None
Match Threshold
40; 43; 46 … 70 … 97%
70%
70%
Auto Collection
On ; Off
On
Off
SVT V. Limit
240; 250; 260 … 650 ms
260 ms
260 ms
Other Enhancements
Stability
Off ; 30; 40 …100 ms
Off
Off
Onset…
Onset
Off ; On; Monitor
Off
Off
Onset Percent
72; 75; 78; 81 ; 84; 88; 91; 94;
97%
81%
81%