7.20.3 Programming considerations for VRS
Auto PVARP and VRS – In the DDIR or DDI mode, when VRS increases the pacing rate,
Auto PVARP reduces the likelihood of competitive atrial pacing.
Mode Switch and VRS – VRS does not operate during Mode Switch episodes.
Conducted AF Response and VRS – In DDIR and VVIR modes, Conducted AF Response
and VRS cannot be programmed to On at the same time.
7.20.4 Programming VRS
Select Params icon
⇒ Pacing…
⇒ Arrhythmia/Post Shock…
▷ V. Rate Stabilization <On>
⇒ Additional V Settings…
▷ Maximum Rate
▷ Interval Increment
7.20.5 Evaluation of VRS performance
The device collects and stores counter data that includes information about the frequency
of PVCs and VRS operation. You can view the stored data on the programmer screen and
print the data.
Select Data icon
⇒ Clinical Diagnostics
⇒ Counters
⇒ VT/VF Episodes
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
293