![Medtronic PROTECTA XT DR D314DRG Скачать руководство пользователя страница 231](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148231.webp)
7.4.1 System solution: Rate Response
Rate-responsive pacing adapts the pacing rate to changes in patients’ physical activity. This
device uses an activity sensor to measure the patient’s movement and to determine the
appropriate pacing rate. It provides dual-slope rate response that may be either automatic
or manual.
7.4.2 Operation of Rate Response
Figure 84. Overview of Rate Response
Activity sensor
Acceleration/
Deceleration
Pacing rate
Rate Profile
Optimization
Rate calculation
The Rate Response system includes an activity sensor to measure patient movement, rate
calculation to convert the patient’s level of physical activity to a pacing rate, Rate Profile
Optimization to automatically adjust rate response settings over time, and acceleration and
deceleration to smooth the pacing rate. This pacing rate is also described as the sensor
rate.
7.4.2.1 Activity sensing
The activity sensor is an accelerometer in the device that detects the patient’s body
movements. Because activity detection varies from patient to patient, the sensitivity to
motion can be adjusted by reprogramming the Activity Threshold parameter. If the Activity
Threshold is lowered, smaller body movements influence the pacing rate. If the Activity
Threshold is raised, body movements must be larger to influence the pacing rate. The
activity count used to calculate the sensor rate is weighted based on the frequency and
amplitude of the accelerometer signal.
The pacing rate is determined by the patient’s level of physical activity and the rate response
parameters. In the absence of activity, such as when the patient is sitting, the pacing rate
is close to the programmed Lower Rate setting. During increased activity, such as when the
patient is walking, the pacing rate is higher.
Medtronic