7.3.2 Operation of MVP mode
Figure 76. Overview of MVP mode
AAI(R)
DDD(R)
Loss of AV conduction
AV conduction resumes
Intact AV
conduction
AV block
7.3.2.1 Intact AV conduction
The MVP modes, AAIR<=>DDDR and AAI<=>DDD, provide AAIR or AAI mode pacing while
monitoring AV conduction. If AV conduction is intact, the device remains in AAIR or AAI
mode. While operating in AAI or AAIR mode, the parameters associated with single
chamber atrial pacing are applicable.
7.3.2.2 Loss of AV conduction
If 2 of the 4 most recent A-A intervals are missing a ventricular event, the device identifies
a loss of AV conduction and switches to the DDDR or DDD mode. The device provides
backup ventricular pacing in response to dropped ventricular events until the loss of AV
conduction is identified.
Figure 77. Switching from AAIR mode to DDDR mode
A
P
A
P
A
P
A
P
A
P
A
P
A
P
A
P
V
P
V
P
V
P
V
P
V
S
V
S
V-V intervals
ECG
Marker Channel
200 ms
1 The device operates in AAIR mode.
2 At the onset of AV block, the device supplies ventricular backup pacing pulses.
3 The device switches to DDDR mode.
Medtronic
PROTECTA™ XT DR D314DRG
224
Clinician Manual