10.5.1 Considerations for performing a Sensing Test
DOO, VOO, and AOO pacing modes – The Sensing Test cannot be performed if the
programmed pacing mode is DOO, VOO, or AOO.
Pacing modes available – The pacing modes available under Test Value depend on the
programmed pacing mode.
Patient comfort – During a Sensing Test, reduce the pacing rate gradually to minimize
patient symptoms associated with abrupt changes in heart rate.
Automatic timeout – The Sensing Test ends automatically after a few seconds and
restores the programmed settings if no intrinsic events occur and no changes are made to
the pacing rate.
Comparison to sensing trends – Sensing amplitude measurements taken during a
Sensing Test may include events that are atypical or a result of oversensing (for example
PVCs or far-field R-waves). These events are excluded from the daily automatic sensing
amplitude measurements the device collects and reports in the sensing amplitude trends.
Because of this difference in measurement operations, Sensing Test results may differ from
those reported in the sensing amplitude trend data.
RV Sense Polarity – The ventricular sensing electrodes included in Sensing Test
measurements depend on the programmed RV Sense Polarity value.
Maximum measured value – The maximum amplitude value that the Sensing Test can
measure is 20 mV. If the amplitude is over 20 mV, the results are displayed as >20 mV.
Selecting sensitivity values – Do not adjust the values for A. Sensitivity and RV Sensitivity
based on the results of the Sensing Test. For more information, see Section 7.1, “Sensing
intrinsic cardiac activity”, page 201.
Tachyarrhythmia detection suspended – Tachyarrhythmia detection is suspended
during the Sensing Test.
10.5.2 How to perform a Sensing Test
Caution: Use caution when selecting temporary pacing settings for pacemaker-dependent
patients. These patients may not receive adequate pacing support while amplitude
measurements are being obtained.
1. Interrogate the device.
2. Select Tests > Sensing.
3. Program the Test Value parameters for Mode and AV Delay or accept the values
displayed.
4. Select [START Measurement].
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
433