Figure 107. Low Rate Episode
7.9 Promoting the intrinsic rate during periods of inactivity
The patient’s intrinsic heart rate is preferable to pacing during extended periods of patient
inactivity, such as when the patient is sleeping.
7.9.1 System solution: Rate Hysteresis
Rate Hysteresis allows intrinsic rhythms to occur below the programmed Lower Rate.
7.9.2 Operation of Rate Hysteresis
Rate Hysteresis is available when the pacing mode is programmed to VVI or AAI, and
functions when the device is operating in one of these modes.
Rate Hysteresis allows a slower lower rate when the intrinsic rate is below the programmed
Lower Rate. After each sensed event, the programmed hysteresis rate is applied. After each
paced event, the programmed Lower Rate is applied.
Medtronic
PROTECTA™ XT DR D314DRG
262
Clinician Manual