7.2.5.2 Programming considerations for dual chamber pacing
SAV and PAV intervals – The SAV interval is usually programmed 30 ms to 50 ms shorter
than the PAV interval. This is done to compensate for the inherent delay between the actual
cardiac event in the atrium and when it is detected by the device.
Upper Tracking Rate – When programming higher upper tracking rates, SAV and PVARP
should be programmed to appropriate values to assure 1:1 tracking (see Section 7.2.8,
“Tracking rapid atrial rates”, page 220).
Upper rates and refractory periods – A combination of a high Upper Sensor Rate and a
long refractory period may result in competitive atrial pacing (see Section 7.2.8, “Tracking
rapid atrial rates”, page 220). Consider programming Non-Competitive Atrial Pacing
(NCAP) to On.
Pacing safety margins – Pacing pulses must be delivered at an adequate safety margin
above the stimulation thresholds.
High pacing output levels – The pulse width and amplitude settings affect the longevity
of the device, particularly if the patient requires bradycardia pacing therapy most of the time.
Cross-chamber sensing – Pulse width and amplitude settings can affect cross-chamber
sensing. If you set the pulse width and amplitude values too high, pacing pulses in one
chamber may be sensed in the other chamber, which could cause inappropriate inhibition
of pacing.
7.2.6 Programming pacing therapies
Select Params icon
⇒ Pacing…
▷ Mode
▷ Lower Rate
▷ Upper Track
▷ Upper Sensor
▷ Atrial Amplitude
▷ Atrial Pulse Width
▷ RV Amplitude
▷ RV Pulse Width
⇒ Paced AV…
▷ Paced AV
▷ Sensed AV
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
219