7. Verify that the device time and date are correct. If necessary, reprogram the time and
date.
8. Perform a manual capacitor formation to reset the capacitor formation timer and to
ensure that the capacitor formation schedule is not compromised.
9. Interrogate the device again. Check the Battery and Lead Measurements screen to
verify that the battery voltage and charge time are acceptable.
10. Conduct lead impedance and pacing threshold tests as desired.
6.15 Optimizing device longevity
Optimizing device longevity is a desirable goal because it may reduce the frequency of
device replacement for patients. Optimizing device longevity requires balancing the benefit
of device therapy and diagnostic features with the energy requirements placed on the
battery as a result of these features.
To view the Recommended Replacement Time (RRT) indicator for the device, refer to the
Quick Look II screen. For information about the longevity of the device, see Section A.3,
“Projected service life”, page 448.
The following sections describe strategies that may help reduce the energy requirements
placed on the battery.
6.15.1 Promoting intrinsic AV conduction
Managed Ventricular Pacing (MVP) – MVP promotes AV conduction by reducing
unnecessary right ventricular pacing. This primary benefit of MVP is therapeutic but it may
also increase device longevity as a result of a decrease in the percentage of pacing. For
more information about MVP, see Section 7.3, “Reducing unnecessary ventricular pacing
with MVP mode”, page 223.
Promoting AV conduction with longer AV intervals – Another method of promoting AV
conduction is to increase the Paced AV and Sensed AV intervals. This allows intrinsic
conduction to occur before a ventricular pace. Fewer pacing pulses may increase device
longevity. For more information, see Section 7.2, “Providing pacing therapies”, page 212.
6.15.2 Managing pacing outputs
Capture Management – Capture Management provides the device with automatic
monitoring and follow-up capabilities for managing pacing thresholds in the right ventricle
and atrium. This feature is designed to monitor the pacing threshold and, optionally, to adjust
the pacing outputs to maintain capture. Programming Capture Management allows the
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
197