Table 7. SessionSync status icon states (continued)
Part of SessionSync sta-
tus icon
Color
What the color indicates
Red circle with a line
through it
A device application in use that does not sup-
port SessionSync
Programmer status
Gray
No session data files in the Transfer Queue
Blue
Session data files in the Transfer Queue
4.14.4.3 How to view the status of the SessionSync feature from the
SessionSync Status screen
The SessionSync Status screen displays information about the data files being transferred
to the data management system using the SessionSync feature. Each status message
includes the date, time, and event information for the associated SessionSync event.
1. From the Desktop, select Programmer > SessionSync Status.
2. Select the [Update Status] button.
Note: Events displayed on the SessionSync Status screen are not automatically updated
when SessionSync Status is selected from the menu. The user must manually select the
[Update Status] button to refresh the events that are displayed.
4.14.5 SessionSync error messages
Table 8. SessionSync error messages
Error Message
What this means
Ending a Session without Automatic
SessionSync
You have deselected the Automatic SessionSync
check box on the End Session window before select-
ing the [End Now] button.
Interrogation Required
You must conduct an interrogation.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
101