4. If you want to treat the induced episode with a manual therapy, select [Suspend] to
prevent automatic detection.
Note: The [Suspend] button is located at the top of the screen and is not shown in
5. Select the Resume at BURST check box for automatic detection and therapy, or clear
the check box for manual therapy.
6. Accept the displayed test values or select new test values.
7. To view and adjust VF detection and therapy parameters, select [Adjust
Permanent…].
8. Press and hold [50 Hz BURST Press and Hold]. Release the button to end the
induction.
9. If necessary, select [ABORT] to abort a therapy in progress.
10. Select [Retrieve Data…] and [Print…] to review and print patient data.
10.7.4 Inducing AT/AF with an atrial 50 Hz Burst
You may use an atrial 50 Hz Burst to induce AT/AF. To induce AT/AF, the 50 Hz Burst
induction delivers a rapid burst of AOO pacing pulses to the atrium. You may specify the
amplitude and pulse width of these pulses but the pacing interval is fixed at 20 ms.
As long as you press and hold the [50 Hz BURST Press and Hold] button on the programmer
screen, the device continues delivering the induction (up to a maximum of 10 s).
If you perform an atrial 50 Hz Burst induction, you may choose to have the device deliver
VOO Backup pacing.
The atrial 50 Hz Burst may also be used to manually treat AF episodes.
Warning: Monitor the patient carefully when using an EP study function. Have an external
defibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmia
may degenerate to ventricular fibrillation.